The Tousimis SAMDRI-PVT-3B is a critical point drying apparatus used after HF-release of a MEMS device to prevent stiction in the device. The apparatus achieves this by avoiding the sudden change in densities caused by a fluid’s transition from liquid to gas during typical air-drying. By using a critical point transition, the transition from liquid to vapor occurs at the same density.
The SAMDRI process begins with immersion of the device in an intermediate dehydrating fluid (100% methanol) in the chamber. The process can then be divided into four stages:
1.
The device and the
chamber are cooled and filled with liquid CO2 and maintained at
08C 6 108C.
2.
The sample in the
alcohol under pressure is purged of the alcohol with cold liquid CO2
until all traces of methanol are removed.
3.
The chamber is heated
to a temperature above the critical temperature of CO2. While the chamber volume is constant, there
is an accompanying rise of pressure to above the critical pressure of CO2,
and the liquid phase is converted to the gaseous phase.
4.
The gaseous CO2
is exhausted at a temperature above 31.18C from the chamber. The specimen is now critically dried and ready
for manipulation.
This
transition of the liquid CO2 at its critical point avoids the distortion
from evaporation boundaries and thermal expansion that leads to stiction problems.
Super
critical point drying is usually performed after release of bulk or surface
micromachined devices to allow movement without stiction. This release is usually performed by immersion
of the device in a beaker of concentrated hydrofluoric acid for a specified
period of time, followed by subsequent immersions in several beakers of DI
water.
The
SAMDRI system requires that the device be completely free of water before
introducing it to the chamber. Water
or oils remaining on the sample during critical point drying will not pass
through the critical point with the transitional fluid. Eventually, the water may evaporate from the device surface and
cause damage or stiction problems. To
avoid this, the device should be subsequently placed in increasing concentrations
of methanol, ending in a 100% methanol of the highest possible purity.
Other
precautions to take to insure dehydration of device:
1.
The methanol should
fill the container containing the device in order to reduce exposure to atmospheric
moisture outside the SAMDRI.
2.
The container should
be tightly stoppered with a lid.
3.
The 100% fluid should
be obtained from a newly-opened bottle of ultra-pure methanol or that it contains
a dehydrant to insure no water contamination.
1.
Instrument power
must be on for at least 5 minutes before use.
2.
There should be at
least 25lbs net weight of liquid CO2 in the supply tank.
Note that the tank should come with 50lbs net weight of liquid CO2
in the supply tank. Each run takes 1-2lbs of liquid CO2. Low CO2 levels could confuse the
logic circuitry of the SAMDRI.
3.
Never overtighten
any connectors unless you notice the escape of gas, and then only gently to
stop escape of gas.
4.
If the unit has not
been used for more than 2-3 days, make a dry run (CO2 only) with
the unit.
5.
Time for purging
may vary. Keep in mind that it may
take up to seven minutes to purge 10 ml of ethanol.
6.
Condensation may
occur on the machine at certain points of the drying procedure. It is necessary that the machine be wiped dry
at these points.
·
Liquid CO2
tank connected through a filter to the SAMDRI unit:
![]() |
·
Inlet and Exhaust
Ports on side of SAMDRI unit:
![]() |
·
Metering Valve Buttons:
Buttons are activated by pushing
in. Buttons will be highlighted when
activated. The metering valve buttons
control the on-off activation of the corresponding valves.
![]() |
·
Metering Valves:
Metering valves have been pre-set
by the Super-user. Valves should not
be adjusted except by an experienced authorized user. Metering valves are opened by turning counter-clockwise.
On-off activation of the valves is provided by metering valve buttons.
![]() |
1.
CAUTION – Use extreme care when handling
METHANOL. It is a FATAL
poison. Small internal doses, prolonged
exposure of the skin to the liquid, or continued inhalation of the vapor can
cause blindness. Causes irritation
to skin, eyes, and respiratory tract. Affects
the central nervous system, gastrointestinal tract, and liver. It is also
FLAMMABLE as a liquid or a vapor.
The following chemical ratings are applicable to methanol:
·
Health Rating: 3
- Severe (Poison)
·
Flammability Rating:
3 - Severe (Flammable)
Further information is available by viewing the MSDS sheet on Methanol in the recording binder next to the machine in the MML cleanroom.
2.
CAUTION: During the drying process, gaseous CARBON DIOXIDE could build up in the work area if not properly ventilated.
Insure that the room is well ventilated and that the cooling exhaust
tube is exhausting out of the cleanroom.
CO2 levels greater than 10% could cause fainting.
If you are feeling light-headed during operation of the unit, seek
fresh air immediately. Prolonged exposure
to high levels of CO2 could prove fatal.
3.
CAUTION: Low liquid CO2 levels in the tank could confuse the logic
circuitry of the SAMDRI unit. Check
CO2 level before each drying run to insure there is an adequate
supply. There must be at least 25lbs
of liquid CO2 in the tank.
4.
Safety valves and
rupture disks have been engineered into the SAMDRI unit to prevent excessive
pressure build-up. Above 1300 psi,
a safety valve will begin to release pressure from the chamber. Also, a mechanical Automatic Pressure Control
System will maintain the chamber pressure less than 1600 psi. If the APCS should fail, a rupture disk will
be activated at 2000 psi at 20°C. In addition, safety thermostats should prevent the temperature from
rising above 49°C.
|
Time Start (mins.) |
Time Stop (mins.) |
Procedure |
|
0 |
5 |
SAMDRI
Unit Warm-up
|
|
5 |
6 |
COOL
Procedure |
|
6 |
8 |
FILL
Procedure |
|
8 |
18 |
PURGE
Procedure |
|
18 |
20 |
Repeat
FILL Procedure (if necessary) |
|
20 |
32 |
HEAT
Procedure |
|
32 |
44 |
BLEED
Procedure |
|
44 |
47 |
VENT
Procedure |
|
47 |
- |
Drying
Complete |
1.
Turn Toussimis unit
on by pressing the Power button on the SAMDRI unit. Power light and chamber light will go on.
Allow SAMDRI to warm up for five minutes before use.
2.
Open sealed chamber
on machine by loosening the 3-knurled nuts. Transfer a sufficient amount of 100% pure methanol
into chamber to cover the entire device. Quickly transfer device from beaker to chamber, minimizing the time
exposed to air.
·
The chamber O-ring
should lie completely in its groove.
·
Never grease or wet
the O-ring with methanol.
3. Place the chamber lid down over the chamber and tighten the three knobs by hand with equal pressure
·
DO NOT OVERTIGHTEN!
4.
Make sure all buttons
except for the power button are off. Open the liquid CO2 tank valve by turning counter-clockwise
two full revolutions until you can verbally hear LCO2 flowing into
the machine.
5.
Begin COOL
procedure. Press ON COOL button. The COOL meter valve should be set to 4 turns
counter-clockwise. Do not adjust the
valve. The temperature as read on
the dial should drop from room temperature (about 30°C) to 0°C in 1 minute. At this point, press COOL button OFF. DO NOT CHANGE THE POSITION OF THE COOL METER
VALVE. The cooling function can be
repeated as required independent of any of the other operations.
6.
Begin FILL
procedure. Press ON FILL button. The FILL meter valve should be set to 4 turns
counter-clockwise. Do not adjust the
valve. Allow the chamber to fill with
liquid CO2 for 2 minutes. A
bubble should travel across the quartz viewing window indicating that the
chamber is full. After two minutes,
press OFF FILL button. DO NOT CHANGE
THE POSITION OF THE FILL METER VALVE.
·
During FILL and PURGE
procedures, keep chamber temperature below 7°C (between +7°C and -30°C). If
temperature increases, press ON COOL button until chamber cools to within
acceptable range.
7. Begin PURGE procedure. Place a polyethylene beaker at the end of the polyethylene exhaust tube connected to the purge-vent-bleed port to collect the ethanol. Press ON PURGE button. PURGE-VENT valve should be set at 2 turns counter-clockwise. Do not adjust the valve. FILL valve should already be set at 4 turns. DO NOT CHANGE POSITION OF THE METER VALVES. Allow PURGE to continue for 10 minutes. At the end of the 10 minutes, check exhaust stream for methanol using the paper test described below.
·
Characteristics visible
to the operator during solvent change:
·
The methanol will
appear as “oily currents” within the chamber.
·
The methanol and
CO2 will appear as a liquid or slurry when exiting the exhaust
hose.
·
Once thoroughly exchanged,
the chamber liquid will be perfectly clear. The exiting material will be cold and white.
·
Test the exhaust
for any alcohol by placing a single piece of paper in the exhaust stream.
Appearance of water marks on the paper indicate that there is still
alcohol present in the exhaust. Continue PURGE for 1 minute past the point
where there are no water marks. Note
purge time on data sheet provided in the recording binder.
8. End PURGE procedure. Press off PURGE button. DO NOT CHANGE ESTABLISHED SETTINGS FOR ALL METERING VALVES.
9.
Repeat FILL
procedure to insure that the chamber is filled with liquid CO2.
10.
Begin HEAT
procedure. Press HEAT button ON. Pressure and Temperature will begin to escalate
automatically. Note the appropriate
pressure and temperature in 1 minute increments on the appropriate data sheet
in the recording binder. The chamber
pressure will increase to around 1300 psi, above the critical point of 1070
psi. The chamber temperature will
increase to between 31°C and 39°C. Wait
four minutes after the chamber temperature and pressure is above the critical
points (31.1°C, 1070 psi). After four minutes, end HEAT procedure by pressing
OFF HEAT button.
·
Critical point passage
should take around 12 minutes.
·
During HEAT mode,
the CO2 may exhibit similar schlieren patterns (“oily currents”)
to those in the PURGE mode. This is
caused by the variations in densities due to the heating. This is not methanol in the chamber.
11.
Begin BLEED
procedure. Push ON BLEED button. The BLEED valve should be set to achieve an
exit flow rate of 7-8 SCFH measured by a flow meter attached to the exhaust
tube at the beginning of the bleed procedure.
Confirm the BLEED valve setting by putting the flow meter into the
polyethylene tube connected to the Purge-Vent-Bleed Port.
If the flow rate is different than 7-8 SCFH, please contact the Super-user.
The pressure should decrease at a rate of about 100 psi per minute. Chamber temperature should remain constant. When the chamber pressure reaches 350 psi,
end BLEED procedure by pushing BLEED button OFF.
·
Decompression should
take around 12 minutes.
12.
Begin VENT
procedure. Push ON VENT button. The VENT-PURGE metering valve has already been
set at 2 turns. The pressure in the
chamber should reach 0 psi in 3-5 minutes.
The flow meter at the start will be 15-20 SCFH. When chamber appears to reach 0 psi, wait 2
more minutes. End VENT procedure by
pushing off VENT button.
13.
The device has now
been critically dried. Open chamber
by loosening the three nuts on the chamber top. Quickly remove sample to a dehydrated clean
environment to prevent further rehydration and stiction.
14.
Make certain the
chamber is clean and the O-ring is in place and in good condition.
Replace lid and tighten by hand. Close
liquid CO2 tank valve by turning clockwise until tight.
15.
Begin Power down
procedure. Press COOL button ON.
Metering valve should already be set at 4 turns.
This will empty the lines of CO2.
Turn OFF COOL button. Do not
adjust or close any of the metering valves.
Turn off POWER button.
16.
Wipe unit down and
cover.
1.
POWER on machine
– (5-minute warm-up).
2.
Place sample in chamber
with methanol.
3.
Open CO2
tank.
4.
COOL – Press on COOL button, meter
at 4 turns (1 minute).
5.
FILL – Press on FILL button, meter
at 4 turns (2 minutes).
6.
Setup beaker for
methanol-LCO2 exhaust.
7.
PURGE – Press on PURGE button, meter
at 2 turns (10 minutes).
8.
Re-FILL – Press on FILL button, meter at
4 turns (2 minutes).
9.
HEAT – Press on HEAT button, record
T & P (12 minutes).
10.
BLEED – Press on BLEED button, meter
set for 7-8 SCFH, record T & P (12 minutes).
11.
VENT – Press on VENT button, meter
at 2 turns (3-5 minutes).
12.
Remove sample from
chamber.
13.
Close CO2
tank.
14.
Exhaust lines – Press
on COOL button, meter at 4 turns (1-2 minutes).
15.
POWER off machine.
16.
Wipe and cover.
Fred Rosenburger
basement@eng.umd.edu
2/7/2000